- Small yet powerful turbopump with a pumping speed of up to 35 l/s for N2
- Integrated TC 110 drive electronics
- For installation in any orientation
- Ideal for dependable systems integration
- Extensive accessories expand the range of applications
Bearing | Hybrid |
Compression ratio for Ar | > 1 · 1011 |
Compression ratio for H2 | 1.4 · 105 |
Compression ratio for He | 1.3 · 107 |
Compression ratio for N2 | > 1 · 1011 |
Connection flange (in) | DN 40 ISO-KF |
Connection flange (out) | DN 16 ISO-KF/G ¼" |
Cooling method | Convection |
Cooling method, optional | Air Water |
Cooling water flow | 75 l/h |
Cooling water flow, max. | 75 l/h |
Cooling water flow, min. | 75 l/h |
Cooling water temperature | 5-25 °C | 41-77 °F | 278-298 K |
Current, max. | 4,6 A |
Electronic drive unit | with TC 110 |
Final pressure | < 1 · 10-7 hPa | < 7.5 · 10-8 Torr | < 1 · 10-7 mbar |
Fore-vacuum max. for N2 | 22 hPa | 16.5 Torr | 22 mbar |
Gas throughput at final rotation speed for Ar | 0.54 hPa·l/s |
Gas throughput at final rotation speed for H2 | 15.3 hPa·l/s |
Gas throughput at final rotation speed for He | 2.7 hPa·l/s |
Gas throughput at final rotation speed for N2 | 1.3 hPa·l/s |
I/O interfaces | RS-485, Remote |
Mounting orientation | Arbitrary |
Operating voltage: V DC | 24 V DC |
Permissible radial magnetic field max. | 3.8 mT |
Power consumption max. | 110 W |
Pumping speed for Ar | 30 l/s |
Pumping speed for H2 | 38 l/s |
Pumping speed for He | 41 l/s |
Pumping speed for N2 | 35 l/s |
Rotation speed ± 2 % | 90,000 rpm | 90,000 min-1 |
Rotation speed variable | 50 – 100 % |
Run-up time | 1.7 min |
Sound pressure level | ≤48 dB(A) |
Venting connection | G 1/8" |
Weight | 2.4 kg | 5.29 lb |