Ambient temperature |
5-55 °C | 41-131 °F | 278-328 K |
Anode |
Mo |
Bakeout temperature |
Electronics removed 150 °C | Electronics removed 302 °F | Electronics removed 423 K |
Feedthrough |
Glass |
Filament |
Tungsten |
Flange, material |
Stainless Steel 1.4435 |
Measurement cable length |
300 m |
Measuring method |
Pirani/Cold Cathode |
Measuring range |
1 · 10-9 – 1 · 103 hPa |
Nominal diameter |
DN 25 ISO-KF |
Output signal: Measuring range |
2 – 8.6 V |
Output signal: Minimum load |
10 kΩ |
Precision: 1 · 10-8 – 1 · 100 hPa |
± 30 % |
Pressure max. |
10,000 hPa | 7,500 Torr | 10,000 mbar |
Repeatability: 1 · 10-8 – 100 hPa |
± 5 % |
Supply: consumption max. |
2 W |
Supply: Voltage V DC |
14.5 – 30 V DC |
Temperature: Storage |
-40-70 °C | -40-158 °F | 233-343 K |
Volume |
19.9 cm³ |
Weight |