OmniStar™ GSD 320 O1, 1-100 amu, tungsten filament, heated capillary 200 °C, 1 m
Overview
Technical data
Analyzer | QMA 200 M |
Contribution to neighboring mass: 40 to 41 | < 10 ppm |
Detection limit min.: C-SEM (ppm) | < 1 ppm |
Detection limit min.: Faraday (ppm) | < 20 ppm |
Gas connection | Stainless steel capillary |
Gas connection: Length | 1 m |
Gas flow rate | 1 – 2 sccm (0 °C) |
Gas inlet | Controlled inlet valve |
Input: Analog | 2 AI, -10-+10 V, 14 bit |
Input: Digital | 1 contact, V in = 24 V DC |
Mass range | 1 – 100 u |
Operating temperature: Capillary, up to | 200 °C | 392 °F | 473 K |
Output: Analog | 2 AO, 0-10 V, 14 bit |
Output: Digital | 2 Relais N.O., 24 V DC |
Power consumption | 590 W |
Quadrupole controller | QME 220 M |
Resolution, adjustable at 10 % peak height | 0.5 - 2.5 u |
Rod system: Diameter | 6 mm | 0.24 inch |
Rod system: Length | 100 mm | 3.94 inch |
Rod system: Material | Stainless steel |
Sample pressure: Standard version | 1,000 hPa | 750 Torr | 1,000 mbar |
Software | Quadera® |
Supply | 100-230 V, 50-60 Hz |
Weight | 35 kg | 77.16 lb |
Accessories
Standard connection OmniStar™ / ThermoStar™
Adapter for heating hose, ¼" (BG 442 778-T)
Dimensions
- A
- 545 mm
- B
- 395 mm
- C
- 305 mm