ASM 1002, 25 m3/h, small test chamber, remote control in mbar l/s, English language, 200/240 V and US mains cable
Order number: Y0R0S10MB810

  • Rotary vane pump with 25 m3/h backing capacity and high vacuum pump 200 l/s
  • 4 l/s helium pumping speed
  • 200/240 V operating voltage
  • US power cable
  • Voice synthesizer in English
  • Small hemispherical test chamber 81 cm3 internal volume (72 mm diameter, 31 mm depth) and automatic cycle start
  • Remote control in mbar l/s
Backing pump with oil sealed backing pump
Backing pump capacity 25 m³/h | 14.71 cfm | 416.67 l/min
Detectable gases 4He
Dimensions (L x W x H) 697 x 867 x 1,244 mm | 27.44 x 34.13 x 48.98 inch
Flange (in) No flange but small test chamber
I/O interfaces Analog outputs (Helium signal log, Inlet pressure); Specific analog outputs to recover the lights status in Pass/Fail
Inlet test pressure, max. 100 hPa  
Language English
Mains cable Yes, US power cable  
Minimum detectable leak rate for helium (sniffing leak detection) 1 · 10-8 Pa m3/s | 7.5 · 10-8 Torr l/s | 1 · 10-7 mbar l/s
Minimum detectable leak rate for helium (vacuum leak detection) 1 · 10-12 Pa m³/s  
Noise level 55 dB(A)
Power consumption < 1500 W
Power consumption max. 2300 W
Pumping speed for He 4 l/s
Sound pressure level 55 dB(A)
Start-up time (20°C) with calibration < 4.5 min
Start-up time (20°C) without calibration < 3.5 min
Supply 200 – 240 V, 50/60 Hz
Test method Vacuum and sniffing leak detection
Type Workstation
User interface Basic user interface
Weight 190 kg | 418.87 lb

Dimension-TIM-RS

File Type Language Date Filesize
Data sheet PDF EN 07-23-2019 ~ 447.5 KB
File Type Language Date Filesize
Operating instructions ASM 1002 (111226OEN_05) PDF EN 08-02-2016 ~ 6.2 MB