QR-Code

PM P04 679
Similar Image

HiPace® 800 with TC 400 and power supply pack OPS 400, DN 200 ISO-F
Order number: PM P04 679

  • Compact yet powerful turbopump with a pumping speed of up to 790 l/s for N2
  • DN 200 ISO-F
  • Drive electronics TC 400 with power supply OPS 400 integrated
  • Ideal for analytical and indurstrial applications
  • For installation in any orientation
  • Extensive accessories expand the range of applications
Bearing Hybrid
Compression ratio for Ar 1 · 1011
Compression ratio for H2 4 · 105
Compression ratio for He 3 · 107
Compression ratio for N2 1 · 1011
Cooling method, optional Air
Cooling method, standard Water
Cooling water flow 100 l/h
Cooling water flow, max 100 l/h
Cooling water flow, min 100 l/h
Cooling water temperature 15-35 °C | 59-95 °F | 288-308 K
Corrosive gas version No
Current max. 8,75 A
Electronic drive unit TC 400
Flange (in) DN 200 ISO-F
Flange (out) DN 25 ISO-KF
Fore-vacuum max. for N2 11 hPa | 8.25 Torr | 11 mbar
Gas throughput at full rotational speed for Ar 3.5 hPa·l/s
Gas throughput at full rotational speed for H2 14 hPa·l/s
Gas throughput at full rotational speed for He 20 hPa·l/s
Gas throughput at full rotational speed for N2 6.5 hPa·l/s
I/O interfaces RS-485, Remote
Integrated power supply YES
Interface, extended Profibus, DeviceNet, E74
Low vibrations No
Mounting orientation Any
Operating voltage: Range 5 %
Operating voltage: V DC 48 (± 5 %) V DC
Permissible magnetic field max. 6 mT
Power consumption max. 420 W
Protection category IP54
Pumping speed for Ar 780 l/s
Pumping speed for H2 580 l/s
Pumping speed for He 700 l/s
Pumping speed for N2 790 l/s
Rotation speed ± 2 % 49,200 rpm | 49,200 min-1
Rotation speed variable 60 – 100 %
Run-up time 2 min
Sound pressure level 50 dB(A)
Ultimate pressure according to PNEUROP 1 · 10-7 hPa | 7.5 · 10-8 Torr | 1 · 10-7 mbar
Ultimate pressure without gas ballast 1 · 10-7 hPa | 7.5 · 10-8 Torr | 1 · 10-7 mbar
Venting connection G 1/8"
Weight 15.4 kg | 33.95 lb
File Type Language Date Filesize
Data sheet PDF EN 10-15-2019 ~ 599.5 KB
File Type Language Date Filesize
Operating instructions HiPace 800 with TC 400 (PT0273BEN_G) PDF EN 06-17-2016 ~ 4.8 MB
Operating instructions TC 400, Electronic drive unit for HiPace (PT0203BEN_T) PDF EN 04-01-2019 ~ 2.2 MB
Operating instructions OPS 400, Integrated Power Supply (PT0416BEN_B) PDF EN 09-10-2014 ~ 1.3 MB
Your local contact
Pfeiffer Vacuum Inc.
24 Trafalgar Square
NH 03063-1988 Nashua
USA

+1 (800) 248-8254
+1 (888) 658-7983
write message