QR-Code

PT M16 112 211
Similar Image

PrismaPro® QMG 250 M1, 1 – 100 u, open ion source, yttriated iridium, I/O-option: Extended
Order number: PT M16 112 211

  • PV MassSpec software for Windows 11 & 10 for recording, saving and displaying measured data
  • Leak detection and vacuum diagnosis with a click
  • Simple definition of measurement recipes
  • Low detection limit
Ambient temperature: analyzer 150 °C | 302 °F | 423 K
Ambient temperature: electronics 5-50 °C | 41-122 °F | 278-323 K
Bakeout temperature: analyzer 300 °C | 572 °F | 573 K
Connection
Connection flange (in) DN 40 CF-F
Connection for gauge Connection for an ActiveLine or DigiLine gauge with analog relay option
Contribution to neighboring mass: 40 to 41 10 ppm
Detection limit min.: C-SEM (hPa) 2.7 · 10-15 hPa | 2.02 · 10-15 Torr | 2.7 · 10-15 mbar
Detector Faraday
C-SEM
Filament Iridium coated with Y2O3
I/O interfaces Ethernet
Input: Analog 1 x ± 10 V / 12 Bit
8 x ± 10 V / 16 Bit
Input: Digital 7 x 24 V
Input voltage(s) 100 – 240 V AC, 50/60 Hz  
Interface: Connection, device side Extended interface, D-Sub, 15 pins, female; HD D-Sub, 62 pins, female
Ion source Open, 2 filaments
Mass range 1 – 100 u
Measuring speed 1 ms/u - 16 s/u
Mounting orientation Arbitrary
Operating pressure max.: Faraday 5 · 10-4 hPa | 3.75 · 10-4 Torr | 5 · 10-4 mbar
Operating pressure max.: SEM 5 · 10-5 hPa | 3.75 · 10-5 Torr | 5 · 10-5 mbar
Output: Analog 8 x 0 – 10 V / 16 bit
Output: Digital 10 x 24 V
Peak ratio reproducibility 0.5 %
Relay: Number 1 pcs.
Resolution, adjustable at 10 % peak height 0.5 – 2.5 u
Rod system: Diameter 6 mm | 0.24 inch
Rod system: Length 125 mm | 4.92 inch
Sensitivity for Ar: Faraday 5 · 10-4 A/hPa
Temperature: Storage -25-70 °C | -13-158 °F | 248-343 K
Weight 3.2 kg | 7.05 lb

Dimension-TIM-RS

Your local contact
Pfeiffer Vacuum Inc.
24 Trafalgar Square
NH 03063-1988 Nashua
USA

+1 (800) 248-8254
+1 (888) 658-7983
write message