Product Selector
Residual gas analysis in ultra high vacuum - Pfeiffer Vacuum

QR-Code

PT M06 111 110
Similar Image

QMG 220 M1, PrismaPlus®, C-SEM, 1-100 amu, open ion source, W, 200 °C, 0°
Order number: PT M06 111 110

  • PrismaPlus® compact mass spectrometer
  • Quadera®-Software for Windows XP Professional and Vista Ultimate
  • Leak detection and residual gas analysis routines launched by a simple mouse-click
  • Determination of gas concentrations by means of integrated matrix calculation
  • Lowest detectable partial pressure: 1 · 10-14 hPa
  • Ethernet networking
Bakeout temperature: analyzer 200 °C | 392 °F | 473 K
Connection
Contribution to neighboring mass: 40 to 41 10 ppm
Detection limit min.: C-SEM (hPa) 1 · 10-14 hPa | 7.5 · 10-15 Torr | 1 · 10-14 mbar
Detection limit min.: Faraday (hPa) 5 · 10-12 hPa | 3.75 · 10-12 Torr | 5 · 10-12 mbar
EMC EN 61326-1
Filament Tungsten
Flange (in) DN 40 CF-F
Input: Digital Ext_Protection, 24 V
Interfaces Ethernet
Ion source Open, 2 filaments
Mass range 1 – 100 u
Measuring channels 128 Pcs.
Mounting orientation Any
Operating pressure max.: C-SEM 1 · 10-5 hPa | 7.5 · 10-6 Torr | 1 · 10-5 mbar
Operating pressure max.: Faraday 1 · 10-4 hPa | 7.5 · 10-5 Torr | 1 · 10-4 mbar
Operating temperature: analyzer 150 °C | 302 °F | 423 K
Operating temperature: electronics 0-40 °C | 32-104 °F | 273-313 K
Peak ratio reproducibility 0.5 %
Resolution, adjustable at 10 % peak height 0.5 - 2.5 u
Rod system: Diameter 6 mm | 0.24 inch
Rod system: Length 100 mm | 3.94 inch
Safety EN 61010-1 / IEC 1010
Scan speed: MID 2 ms - 60 s/amu
Scan speed: scan analog 20 ms - 60 s/amu
Scan speed: scan bargraph 20 ms - 60 s/amu
Sensitivity for Ar: C-SEM 200 A/hPa
Sensitivity for Ar: Faraday 5 · 10-4 A/hPa
Temperature: Storage -25-70 °C | -13-158 °F | 248-343 K
Weight 3.8 kg | 8.38 lb

Massbild-TIM-RS

Accessory for PrismaPlus QMG 220

IO 220, I/O-Modul for PrismaPlus® PT M28 699
File Type Language Date Filesize
Data sheet PDF EN 06-22-2018 ~ 359.1 KB
Stepfile STP EN 01-05-2017 ~ 8.9 MB
File Type Language Date Filesize
Operating instructions QMG 220 PrismaPlus® - Compact Mass Spectrometer (Communication Protocol) (BG5215BE) PDF EN 08-30-2017 ~ 1.8 MB
Operating instructions QMG 220 PrismaPlus® - Compact Mass Spectrometer (BG5214BEN_D) PDF EN 06-06-2017 ~ 3 MB
Software LabView Installation instructions (PrismaPlus) (PG0036) PDF EN 01-09-2018 ~ 1.7 MB
Software LabView Driver - PrismaPlus ZIP EN 01-09-2018 ~ 7.1 MB
Your local contact
Pfeiffer Vacuum Inc.
24 Trafalgar Square
NH 03063-1988 Nashua
USA

+1 (800) 248-8254
+1 (888) 658-7983
write message